D-Tec Vaporizer
Page 21
O & M Manual Part No. 1107-0622-000
August 2001
Alarm Indication
Action
Possible Fault
Fill the sump with
Suprane™.
Change the battery within
eight hours of alarm
indication
w
Warning: Do not
attempt to use the
vaporizer.
Agent level < 60 ml.
w
Warning: If this
condition persists
until there is less
than 20 ml of agent
in the sump, the NO
OUTPUT alarm is
activated and the
output ceases.
Alarm battery low.
If the amber WARM-UP light
does not extinguish within
ten minutes after the
vaporizer is connected to an
electrical supply, there may
be an internal malfunction.