D-Tec Vaporizer
Page 17
O & M Manual Part No. 1107-0622-000
August 2001
DURACELL
DU
RA
CE
LL
mn
6
23
r
6.2
Changing The Battery
The battery must be changed every 12 months irrespective of condition.
1.
Turn the dial to
q
and disconnect the mains lead from the electrical supply.
2.
Unscrew the battery cover securing screw and remove the battery cover from the base of the
vaporizer.
3.
Remove the battery from the vaporizer and disconnect the terminals. Dispose of the battery.
w
Caution: Only a Duracell 1604 or VARTA Energy 2000, 9 Volt battery must be installed in the vaporizer.
If any other battery is installed, it may damage the vaporizer.
w
Warning: The battery terminals must be firmly clipped onto the battery to help prevent a possible
disconnection when the vaporizer is moved.
4.
Clip the terminals firmly onto a new battery observing the correct polarity and insert the battery into
the vaporizer. Fit the battery cover to the base of the vaporizer and tighten the securing screw.
BATTERY COVER
SECURING
SCREW
BATTERY
TERMINALS
DURACELL 1604 OR VARTA
ENERGY 2000 BATTERY
AB.09.007
Fig. 18 Changing The Battery
w
Warning: Do not throw batteries onto a fire, do not open with force and do not recharge. To do so may
result in operator injury. When disposing of batteries, they must be treated as special waste and
disposed of according to local regulations.
6.3
Spare Parts
Description
Part No.
Mains Lead, North America - Long* 1107-3189-000
Mains Lead, North America - Short 1107-3190-000
Battery
1373256
* All units ship with a long cord