Aespire View
10-32
M1132382
System components
Integral
This anesthesia system contains the following integral components,
monitoring devices, alarm systems, and protection devices that
comply with European, international, and national standards:
•
Breathing system pressure-measuring device.
•
Airway pressure-limitation device.
•
Exhaled-volume monitor.
•
Breathing system integrity alarm.
•
Breathing system continuing-pressure alarm.
•
O2 monitor.
•
Anesthesia ventilator.
•
Breathing system.
Not integral
These devices are not integral to this anesthesia system:
•
CO2 monitor.
•
Anesthetic agent monitor.
WARNING
European, international, and national standards require
the following monitoring be used with this system:
•
Exhaled volume monitoring.
•
O2 monitoring.
•
CO2 monitoring.
•
Anesthetic agent monitoring be used when anesthetic
vaporizers are in use.
When adding devices to the anesthesia systems, follow the
installation instructions provided by the device manufacturer.
Whoever adds individual devices to the anesthesia system shall
provide instructions on how to enable the individual devices (for
example, a preoperative checklist).