3 ½   D I G I T ,   L E D   D I S P L A Y   D I G I T A L   P A N E L   V O L T M E T E R S

DMS-30PC

Figure 10. External Gain Adjustment

Figure 8. 4.5-18V Power Supply Monitor

Figure 9. Basic DC Ammeter Circuit

Figure 7. 4-to-20mA Current Loop Operation

To calibrate the circuit of Figure 7, perform the following:

1. With 4mA applied, adjust the 50k

 potentiometer (R2) to

display a reading of "000" (assuming that is the desired
reading).

2. With 20mA applied, adjust the gain-adjust potentiometer on

the back of the meter to display a reading of "1000".  For
different full scale readings, alter the value of R

Shunt

 accordingly.

Example: For a meter with a 2V full scale input (1.999 full
scale reading) and a desired display reading of
"1000" (with an input of 20mA),  V

Fsr

 

= 1.000 Volts

R

Shunt

= 1.000V/(0.020 – 0.004)A

R

Shunt

= 1.000V/0.016A = 62.5 Ohms

6Power Supply Monitoring:  One of the most common digital

panel meter applications involves monitoring the output voltage of
the system power supply — often this supply also powers the
meter itself. The low-power, red LED DMS-30PC-2-RL can be
configured to allow power supply monitoring over the range of
4.5-18Vdc. The circuit in Figure 8 uses a low-drop-out, three-
terminal regulator (LM-2931Z-5, available from National
Semiconductor) to provide regulated 5V power to the meter. The
LM-2931 was chosen because it has the following on-chip

protection features: reverse polarity, short circuit and thermal run-
away. When using other, higher-power, DMS-30PC models with
three-terminal regulators, be sure to consult the regulator
manufacturer's data sheet to ensure the regulator is being
utilized safely and correctly.

7. Digital Ammeter:  Digital ammeters are finding ever-increasing

usage because analog-style ammeters (moving-vane types) now
cost roughly the same as their digital counterparts. Additionally,
analog ammeters are not nearly as rugged as modern digital
panel voltmeters. Figure 9 illustrates a typical ammeter
application. The circuit uses a ±200mV input meter
— the preferred range for most ammeters — to measure the
voltage developed across a 0.1

 current shunt. The circuit

shown represents a basic ammeter connection diagram. A
detailed application note describing digital dc ammeters is
included in DATEL's new Digital Panel Meter Databook.

8. External Gain Adjustment:  Connect REFERENCE OUT

(pin 8) to REFERENCE IN (pin 7) for normal, factory calibrated,
operation.  Use the +1.23V REFERENCE OUT (pin 9) for
applications needing external gain adjustment.  Figure 10 shows
the wiring configuration for each model.  Calibration is performed
with a precise, near-full-scale, input voltage.

Applications

1

3

5V RET

DATEL

UPA-5/500

120 VAC

AC to DC Converter

DMS-30PC-1

12

(–) IN LO

10

ANA COMM

11

(+) IN HI

R1

+

4-20mA

50k

R2

8

7

REF OUT

REF IN

9

1.23V
REF

5V SUP

DMS-30PC-2-RL

5

7

12

+

11

1

4.5 - 18Vdc

LM2931Z-5

GND

IN

OUT

+5V SUP

3

5V RET

DP2

REF

IN

(+) IN HI

8

REF

OUT

(–) IN LO

22µF

10V

+

DMS-30PC-0-YS

6

3

+

1

1.5 Amp (Load Current)

5Vdc

7

8

+5V SUP

5V RET

DP1

REF OUT

REF IN

11

0.150V

(+) IN HI

0.1

R

Shunt

Load

(3.3   )

12

(–) IN LO

9

(+) IN HI

(–) IN LO

ANA COMM

10

1.23V REF

1k

10 to 20 Turns

8

REF IN

11

1k

10 to 20 Turns

8.45k, 1%

8.45k, 1%

5V RET

Connections

for ±2V, ±20V 

and ±200V models

Connections

for ±200mV

models

7

NC

12

3

DMS-30PC

12

VOLTAGE

CALIBRATOR

OUT

COM

Summary of Contents for DMS-30PC Series

Page 1: ...reference and is fully calibrated prior to potting We guarantee outstanding initial accuracy 1 count and excellent stability 0 15 counts C All models operate from a single 5V supply and the low power models draw as little as 10mA 50mW total power A DISPLAY TEST function is standard on each device For popular applications 4 to 20mA rms to dc conversion ac line power J and K thermocouples etc the DM...

Page 2: ... for 2V models DMS EB TCK K type thermocouple inputs for 2V models DMS EB RMS For true rms measurements of ac voltages DMS EB AC DC For ac line powered applications DMS EB LP For 4 to 20mA loop powered applications Input Range 0 200mV 1 2V 2 20V 3 200V Available on the following models only GS Standard Green RS Standard Red RL Low Power Red Ordering Information Applies for transient or continuous ...

Page 3: ... The multimeter s negative lead usually the black one must be connected to pin 3 5V RETURN 4 DISPLAYTEST Pin 2 Connecting pin 2 to 5V SUPPLY pin 1 will activate all LED segments except the decimal points and the display will read 1888 regardless of the actual applied input If a negative input is applied DISPLAY TEST will also activate the minus sign To protect the LED s the display should not be l...

Page 4: ...s 2 Differential Input Configurations Differential measurements can be made with all DMS 30PC meters Figure 3 though not a practical real world application uses a voltage divider to demonstrate the concept of a differential input signal Be careful not to exceed the 2V common mode voltage limitation for 5V powered meters 3 Engineering Scaling For measuring voltages greater than the full scale input...

Page 5: ...onsult the regulator manufacturer s data sheet to ensure the regulator is being utilized safely and correctly 7 Digital Ammeter Digital ammeters are finding ever increasing usage because analog style ammeters moving vane types now cost roughly the same as their digital counterparts Additionally analog ammeters are not nearly as rugged as modern digital panel voltmeters Figure 9 illustrates a typic...

Page 6: ... 0 025 0 635 NOMINAL RECOMMENDED PC BOARD FINISHED HOLE DIAMETER 0 042 0 003 1 067 0 076 Mechanical Specifications DS 0258D 1 98 MADE IN USA BEZEL INSTALLATION PANEL BEZEL RETAINING CLIP INSTALLATION PANEL 2 56 INSERT 0 156 3 96 DEEP FRONT VIEW 1 270 32 26 0 187 4 75 OPTIONAL BEZEL DMS BZL1 and DMS BZL2 2 35 59 69 2 118 53 80 0 093 2 362 DIAMETER 4 REQUIRED ONLY WHEN USING OPTIONAL BEZEL ASSEMBLY ...

Reviews: