3-5
DATASONICS
User’s Manual
March 1997
3.3 Configuration
There are seven switches which determine the mode of operation. The
different combinations are listed in the table below.
DIP SWITCHES - All dip switch changes will take affect immediately, except
for the baud rate. To change baud rate the unit must be reset(i.e. cycle pwr).
3.4 Predeployment
Always test the altimeter immediately before deployment to ensure that the
unit can transmit and receive acoustic signals and that the system
parameters are correctly set.
Description
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
Test mode - see section 4.4
ON
ON
External key - 200 MS min. between keys
ON
OFF
Internal key - serial out every depth cycle
OFF
ON
Internal key - serial out every fifth depth cycle
OFF
OFF
More gain
ON
Less gain
OFF
Disable under-range detection
ON
Under-range detection (if range is less than
0.8, then output is ’0.8<’)
OFF
1200 baud (rep rate may be > 200ms)
ON
ON
2400 baud
ON
OFF
4800 baud
OFF
ON
9600 baud
OFF
OFF
RS-232 off
ON
RS-232 on
OFF
FSI mode disabled
OFF
FSI mode enabled
ON