![Dataq DI-2108 User Manual Download Page 26](http://html1.mh-extra.com/html/dataq/di-2108/di-2108_user-manual_3691315026.webp)
DI-2108 Hardware Manual
Controls, Indicators, and Connections
20
Acquisition menu to open the Channel Selection grid. Channel 11 is the designated Count channel. Click on the
Channel 11 channel box to enable/disable Counts. The Channel 11 channel box will display a “C” when enabled.
The counter channel accumulates counts over time whether recording or in standby mode as soon as the Counter
channel is enabled. The
accumulated
values will be recorded to your data file unless you reset the counts to zero. The
maximum count value is 65,535 before the counter resets to zero automatically.
Use the Edit >
Reset Count
command in W
IN
D
AQ
Acquisition software to reset counts immediately to zero. This
feature is available in both standby and record modes.
Use the Edit > Preferences >
Reset Count on New File Record
in W
IN
D
AQ
Acquisition software to reset the count at
the start of a new file or when using the Edit > Preferences > Record Next File on Full or Open Next File on Full fea-
ture to reset counts to zero every time a new file is started.
Use the Edit > Preferences >
Reset Count on Manual Record
in W
IN
D
AQ
Acquisition software to reset counts to
zero every time you begin or resume recording a data file using the File > Record command (F4).