RAPID CONFIGURATION
5
1
(Fixed Lens Models only)
The Matrix 300™ fixed lens models are factory focused to a precise Reading Distance. If this
distance is compatible with your application you can use the X-
PRESS™ Interface to install
the reader, if not, use the VisiSet™ procedure described in step 6B.
1. Place the reader at the factory focused Reading Distance and go to step 4.
Factory focus for 6 mm fixed lens models is
85 mm
.
STEP 4
– X-PRESS™ CONFIGURATION
Once Matrix 300™ has focused at the correct reading distance, you must configure it for
optimal code reading relative to your application. This configuration can be performed either
through the X-
PRESS™ Interface or the VisiSet™ configuration program.
Aim
1. Enter the Aim function by pressing and holding the
X-
PRESS™ push button until the Aim LED is on.
2. Release the button to enter the Aim function. The
laser pointers turn on.
3.
Select a single code from your application
.
Position the code at the center of the FOV
(equidistant from the laser pointers).
For Liquid Lens models the reader may start flashing and try
to perform autofocus however this will have no effect on the
application specific code; it can be ignored.
Exit the Aim function by pressing the X-
PRESS™
push button once. After a short delay the Aim
function is cancelled and the laser pointers turn
off.
Figure 5
– X-PRESS™ Interface: Aim
Function
Setup
4. Enter the Setup function by pressing and holding
the X-
PRESS™ push button until the Setup LED is
on.
5. Release the button to enter the Setup function.
The Setup LED will blink until the procedure is
completed.
The Setup procedure ends when the Image
Acquisition parameters are successfully saved in
the reader memory, the Setup LED will stop
blinking
and Matrix 300™ emits 3 high pitched
beeps.
If the calibration cannot be reached after a timeout
of about 5 (five) seconds Matrix 300™ will exit
without saving the parameters to memory, the
Setup LED will stop blinking and in this case Matrix
300™ emits a long low pitched beep.
Figure 6
– X-PRESS™ Interface: Setup
Function