Device adjustment
Should Datacolor 45IR do not work properly, first check, if all instrument settings and measurement
conditions are set correctly. Especially the settings of the
Density adjustment (Slope)
and the
absolute white
reference values
influence the measurement result.
Setting Density SLOPE
With the SLOPE correction values it is possible to adjust Datacolor 45IR to the density values of other devices.
Activate the “soft key” with the cogwheel icon and select the menu
item
Density adjustment (Slope)
.
Select the menu item
Cyan
.
Enter the Cyan value, you find on the calibration chart. By pressing the
arrow buttons the value can be increased or decreased.
In the same way enter the data given by the calibration chart for
Magenta, Yellow and Black.
Select the menu item
Make Slope calibration
.