3 Installation Procedure
l
In the
System Configuration
dialog press the
Add
button
①
to invoke the
Available Control Modules
dialog.
l
Select the
Shimadzu GC- 201x
in the
GC
section and press the
Add
button.
l
The
Shimadzu GC-201x Setup
dialog will appear.
Fig 4: Shimadzu GC-201x Setup
l
Fill in the
Port
,
Baud Rate
and
Name
according to the settings and the
connection of the GC (for more details see the chapter
Shimadzu GC-
201x Setup
on pg
24
)..
l
Press the
Autodetect
button to read all settings from the instrument and
apply them in the
Shimadzu GC-201x Setup
dialog.
Note:
It is strongly recommended not to change the settings gained through the
Autodetect
function. When opening the
Instrument
window, those
settings are checked against the settings of the GC and in case they are
not the same the warning message is issued and no control is possible.
Note:
Only
GC2010
and
GC2014
models are autodetected properly. When
using
GC2014C
or
GC2014 APC/AFC
, the exact type used needs to be
selected after autodetection.
l
Switch to the
Signals
tab (for more details see the
Shimadzu GC- 201x
Setup
dialog section later in the manual ), select the correct number of
signals you want to acquire, name them and possibly check the
This
Device Starts the Run in Clarity
/
Clarity Starts This Device
radiobutton.
The
Clarity Starts This Device
option should be checked only when the
GC will not start the acquisition (that means it will not be triggered by the
GC Start button, external start contact or autosampler). Press the
OK
button.
- 5 -