Wiring Signals
35
shows how to connect a thermocouple input to a DT9828 module.
Figure 6: Connecting Thermocouple Inputs (Shown for Analog Input Channel 0)
Thermocouple
Input
+
-
1
One cold-junction compensation (CJC) circuit is
provided for all thermocouple input channels. The
software reads the value of the CJC input and
applies it to the value of the thermocouple input to
correct for errors based on the specified
thermocouple type.
The output of the channel goes to positive, full-scale
if an open circuit is detected at the input.
2
DT9828
Analog Input Terminal Block