Wiring Signals to the Module
39
Connecting Counter/Timer Signals
DT9812, DT9813, and DT9814 Series modules provide one counter/timer that you can use for
the following operations:
• Event counting
• Frequency measurement
• Edge-to-edge measurement
• Continuous pulse output (rate generation)
This section describes how to connect counter/timer signals for these operation modes. Refer
to
for more information about using the counter/timers.
Event Counting
shows how to connect counter/timer signals to the module to perform an event
counting operation using an external gate.
In this example, the counter counts the number of rising edges that occur on the Counter 0 In
signal when the Counter 0 Gate signal is in the active state (as specified by software). Refer to
for more information.
Figure 12: Connecting Counter/Timer Signals for an Event Counting Operation
Using an External Gate
shows how to connect counter/timer signals to the module to perform an event
counting operation without using a gate (also called a software gate). The counter counts the
number of rising edges that occur on the Counter 0 In signal.
Signal
Source
Ground
Counter 0 In
Ground
External
Gating
Switch
18
16
Counter
0 Gate
19
DT9812, DT9813, or
DT9814 Series Module
Summary of Contents for DT9812 Series
Page 1: ...DT9812 UM 20769 R DT9813 and User s Manual DT9814 Series Title Page ...
Page 4: ......
Page 12: ...About this Manual 12 ...
Page 13: ...13 1 Overview Key Hardware Features 14 Supported Software 15 Getting Started Procedure 16 ...
Page 17: ...Part 1 Getting Started ...
Page 18: ......
Page 26: ...Chapter 2 26 ...
Page 53: ...Part 2 Using Your Module ...
Page 54: ......
Page 76: ...Chapter 5 76 ...
Page 88: ...Chapter 6 88 ...
Page 94: ...Chapter 7 94 ...
Page 113: ...113 B Screw Terminal and Connector Pin Assignments ...