background image

 

Page 12 of 22 

4.2  Rail mounted version (X200LR) 

 

 

Remove the light from its packaging. 

 

If  you  have  also  purchased  our  XRAIL  accessory,  hold  the  rail  against  the 
selected location on the wall, bearing in mind the final position and range of 
travel of the light-head required to project its light patch where it will be 
needed. 

 

Ensuring the rail is horizontal; mark the positions for the 2 screw holes. 

 

Fix the rail to the wall, using appropriate plugs as necessary and the 2 screws 
provided.  

 

NOTE: Slide the screw-head cap retainers the correct way up onto the screws 
before fitting the screws! 

 

 

Fit the screw-head caps. 

 

Ensure the rail end caps are properly fitted. 

 

Slacken sufficiently the light clamp to allow the clamp to pass over the rail 
and fit the light to the rail. 

 

The light can be moved along the rail by slackening the clamp slightly and not 
sufficiently to allow the light to become detached from the rail. 

 

4.3  Desk mounted versions (X200LD / X200LDS / X200LFDM) 

 

X200LD (includes PIVOTD2 desk clamp) 

 

Remove the light from its packaging.  

 

Slacken the desk clamp so that its jaws just clear the desk edge.  

 

Tighten the clamp sufficiently to provide a rigid mounting for the light.  

 

Fit the spigot on the end of the light  arm into the top of the clamp, make 
sure it is fully secured in place. 

 

Apply the power cable, and power the light on. See page 8. 

 
X200LDS (includes PIVOTS weighted base) 

 

Remove the light from its packaging.  

 

Fit the spigot on the end of the light arm into the top of the weighed base, 
make sure it is fully secured in place. 

 

Apply the power cable, and power the light on. See page 8. 

 
X200LFDM (includes PIVOTFDM fixed desk mount)
 

 

Remove the light from its packaging.  

 

Mark the position on the desk for mounting. Use screws the fix the bracket 
securely. 

Summary of Contents for X200

Page 1: ...X200 LED Examination Light Operating Installation Manual QAM X200 0519 12...

Page 2: ...Page 1 of 22...

Page 3: ...4 1 1 Considerations 9 4 1 2 Placement and Installation 10 4 2 Rail mounted version X200LR 12 4 3 Desk mounted versions X200LD X200LDS X200LFDM 12 4 4 Mobile version X200LM 13 4 5 Optional battery ba...

Page 4: ...n DARAY s BioProtect coating an antimicrobial protection against infection causing bacteria such as MRSA The light comprises 3 parts light head and flexible arm system power supply module and the moun...

Page 5: ...ured in Great Britain by Daray Ltd Edison House Robian Way Swadlincote Derbyshire DE11 9DH 1 1 Supported Lights The Following Models are supported in this Manual X200LW X200 LED Wall Mounted Light X20...

Page 6: ...500 Colour Rendering Index CRI 95 Colour Rendering R9 88 Light Field Diameter D50 0 5m 120mm 0 75m 135mm 1 0m 180mm Light Field Diameter D10 0 5m 230mm 0 75m 325mm 1 0m 460mm Number of LED s 3 Power...

Page 7: ...8m 1 Desk Clamp Fixed Desk Mount Desk Stand 1 Wall bracket fixings 1 Light head and flexible arm 1 Thumb Screw for Wall Mount 1 Microfibre cleaning cloth 1 Microfibre cleaning cloth 1 User Manual 1 M...

Page 8: ...Page 7 of 22 3 Operation 3 1 Range of Motion This is an image depicting the range of motion on the X200 light all versions...

Page 9: ...and hold to adjust the intensity It takes approximately 4 seconds to switch between the highest and lowest settings Release the button when at a comfortable setting for the user Press once to power OF...

Page 10: ...to the underside of the bracket dependent on couch height It may be necessary to adjust this height if there are specific medical procedures the light will be used for If in doubt consult the end use...

Page 11: ...in mind the final position of the light head required to project its light patch where it will be needed We recommend 850 1100mm from the floor to the underside of the bracket dependent on couch heig...

Page 12: ...light on See page 8 Check arm and head for smooth and correct movement Your new DARAY light is now ready for use To gain full support of our warranty please fill in our warranty web form or contact DA...

Page 13: ...can be moved along the rail by slackening the clamp slightly and not sufficiently to allow the light to become detached from the rail 4 3 Desk mounted versions X200LD X200LDS X200LFDM X200LD includes...

Page 14: ...g If not already fitted fit the castors to the base by knocking them sharply into their fixing sockets Insert the up stand into the castor base as shown below Do not use a metal hammer Once fitted the...

Page 15: ...e into the black power supply attached to the base upstand The light can now be tested by connecting the DC jacks between the power supply and the light and applying power Press the switch on the head...

Page 16: ...mobile version of this light please use the extension cable provided NOTE Charging time can take up to 8 hours depending on how depleted the battery is 2 If you need to use the light on battery power...

Page 17: ...and that there is no visible cable wear All fixings to be checked and be of sound construction 5 2 Safety Precautions For all cleaning work power off the equipment and where possible unplug the mains...

Page 18: ...our BioProtect antimicrobial coating on painted finishes This is designed to assist with infection control and is not a replacement for cleaning CAUTION Our products are NOT waterproof The equipment...

Page 19: ...ight 5 4 2 Detaching Parts De construction of the lights should be done following a reverse order of the instructions in section 4 The X200 is a completely enclosed light and arm it should not be open...

Page 20: ...e extension cable are fully connected For the mobile version ensure that the IEC is firmly pushed into the PSU Poor Light output Power Button Please check the variable intensity is not just set to its...

Page 21: ...SPARE PARTS or TECHNICAL ISSUES phone DARAY S Service Delivery Team on 44 0 800 878 9864 Opening times 9am 5pm Monday to Thursday excluding bank holidays 9am 2 30pm on Friday Or email support daray co...

Page 22: ...ges may be applicable OTHER Any situation which is not covered by any of the above We will always try to help but we cannot normally offer a refund For additional clarification please refer to our ter...

Page 23: ...Daray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from...

Page 24: ...Page 23 of 22 Service Notes...

Page 25: ...Page 24 of 22...

Page 26: ...Page 25 of 22...

Page 27: ...Page 26 of 22...

Page 28: ...DARAY Ltd Edison House Robian Way Swadlincote Derbyshire United Kingdom DE11 9DH 44 0 800 878 9864 44 0 333 321 0973 sales daray co uk www daray co uk...

Reviews: