Appendix A
Specifications
DAQCard-1200 User Manual
A-8
ni.com
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth .....................................30 VAC or 60 VDC,
installation category I
Channel-to-channel.................................30 VAC or 60 VDC,
installation category I
Environmental
Operating temperature ............................0 to 50 °C
Storage temperature ................................–55 to 150 °C
Humidity .................................................5% to 90% RH, noncondensing
Maximum altitude...................................2000 meters
Pollution degree (indoor use only) .........2
Safety
The DAQCard-1200 meets the requirements of the following standards for
safety and electrical equipment for measurement, control, and laboratory
use:
•
EN 61010-1:1993/A2:1995, IEC 61010-1:1990/A2:1995
•
UL 3111-1:1994
•
CAN/CSA c22.2 no. 1010.1:1992/A2:1997
Electromagnetic Compatibility
CE, C-Tick, and FCC Part 15 (Class A) Compliant
Electrical emissions ................................EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Electrical immunity ................................Evaluated to EN 61326:1997/
A1:1998, Table 1
Note
For full EMC compliance, you must operate this device with shielded cabling.
In addition, all covers and filler panels must be installed. Refer to the Declaration of
Conformity (DoC) for this product for any additional regulatory compliance information.