background image

AN271156466186en-010201 

© Danfoss | Energy Meters | 2020.11 | 8

6. Disposal

This symbol on the product indicates that it will not be treated as household waste. It 
must be handed over to the applicable take-back scheme for the recycling of electrical 
and electronic equipment. For more detailed information about the recycling of this 
product, please contact your local municipal office.

Item

Material

Disposal

Battery

AA-cell Lithium/thionyl chloride 620 mg 
Lithium

Approved deposit for lithium 
batteries

PCBA with display and 
communication module

Coppered epoxy laminate components 
soldered on, PC, TPE

Electronic waste

Cables

Copper with PUR or PVC jackets

Cable recovery

Flow sensor
(including transducer 
and liner)

Brass, stainless steel, PPS

Metal recovery

Transducer

PZT, stainless steel, PEI

Approved deposit for PZT

Other plastic parts

PC, PPS, PEI, TPE

Plastic recovery

7. Local Importer name and address

For goods delivered to UK, importer name and address is:

Danfoss Ltd.
Oxford Road
UB9 4LH Denham
UK

Summary of Contents for SonoSensor 30

Page 1: ...Installation User Guide SonoSensor 30 Ultrasonic flow meter for heating and cooling applications...

Page 2: ...2 Danfoss Energy Meters 2020 11 AN271156466186en 010201 Installation User Guide SonoSensor 30 Note To ensure latest version of declaration please visit danfoss com...

Page 3: ...reparation 4 2 2 Installation of flow sensor 4 2 3 Mounting of electronic unit 5 3 Electrical wiring 6 3 1 Flow pulse cable connection 6 3 2 External power supply module if used 6 3 3 Wiring SonoSenso...

Page 4: ...he flow sensor is installed in the supply or return pipe of the system The flow sensor must be installed so that the direction of flow corresponds to the direction of the arrow on the flow sensor Calm...

Page 5: ...ectronic unit Direct mounting on ultrasonic flow sensor housing turning every 90 C only when the temperature of the flow does not exceed 90 C a On the flow sensor with a thread connection Wall mountin...

Page 6: ...module if used Terminal Nr Description 54 Mains supply 24 V AC DC bipolar 55 Mains supply 24 V AC DC bipolar 3 3 Wiring SonoSensor 30 to energy calculator Infocal 9 SonoSensor 30 in return pipe Infoca...

Page 7: ...sticker on the cover protecting electronic module manufacturer security seal 3 Mounting seal after installation 5 Power supply SonoSensor 30 can be powered from lithium battery 12 years life time or f...

Page 8: ...osal Battery AA cell Lithium thionyl chloride 620mg Lithium Approved deposit for lithium batteries PCBA with display and communication module Coppered epoxy laminate components soldered on PC TPE Elec...