background image

User Guide | Gas detecting sensor, type PAGM for CO2 and Halocarbon

© Danfoss | DCS (vt) | 2021.03

BC372016561395en-000101 | 3

1.  Visual inspection

2.  Connecting gas-sample 

probe and exhaust and 

zero air filter assembly

2.1 

Gas-Sample Probe

2.2 

Purge/Zero Air Carbon 

Filter Assembly

PAGM-IR Halogen gas

 (code no. 080Z2149)

Item

Qty.

Refrigerant monitor

1

Soft carrying case with shoulder strap

1

Battery pack and charger kit

1

Probe assembly

1

Coalescing filter

1

Carbon filter

1

PAGM-IR CO2 

(code no. 080Z2148)

Item

Qty.

Refrigerant monitor

1

Soft carrying case with shoulder strap

1

Battery pack and charger kit

1

Probe assembly

1

Coalescing filter

1

 Hazardous area warning:

This instrument has not been designed to be intrinsically safe for use in areas classified as hazardous 

locations. For your safety, DO NOT use it in hazardous (classified) locations.

Do not operate this equipment in the presence of flammable liquids, vapours or aerosols.

 Combustible/flammable gas warning:

This is NOT a safety device. Some gases which this instrument can detect may be combustible/

flammable. When properly configured, this instrument is designed to alarm at concentrations that 

are lower than the explosive limit of the gas. As such, it is the buyer’s responsibility to initiate an 

immediate planned response to any gas leaks as soon as they are detected. This equipment should 

NEVER be used to measure or sample gases at or above their respective lower explosive limits.

Operation of any electrical equipment in such an environment constitutes a safety hazard.

 Warning:

A gas monitor is a delicate and complex device and it is not to be used in any application that is 

beyond its intended purpose or beyond the scope of its specifications. For details on appropriate use, 

refer to the general description, application, and operation discussions in this manual. Before risking 

equipment damage or personal injury, contact the manufacturer if you are unsure of the validity of a 

particular gas monitor application.

 Warning:

A potential risk exists if the operating instructions are not followed.

 Warning:

To avoid risk of injury from electric shock, do not open the enclosure when power is applied. 

 Caution:

Except for the maintenance detailed in this manual, this product should ONLY be opened and / or 

serviced by authorized personnel. Failure to comply may void the warranty.

When you receive the instrument, please unpack from it from shipping carton and perform visual 

check on the enclosure to ensure no damage to the external parts of the unit which might have 

occurred during shipment. Check that all the below items have been received.

Before first power up connect all necessary accessories as described on below, this will ensure correct 

functionality of the unit. 

To connect the gas-sample probe to the monitor, connect the probe to the gas sample port (see 

picture below).

A Purge/Zero Air Carbon Filter must be installed to properly use the Portable Aspirated Refrigerant 

Monitor (PAGM).

Filter installation:

1.  Unfasten the Velcro on the front of the monitor soft case.

2.  Slide the Purge/Zero Air Carbon Filter into the front pocket. Route the tubing through the hole in 

the back of the pocket, along the side of the housing.

3.  Insert the connector extending from the Purge/Zero Air Carbon Filter into the Zero Air Inlet Port on 

the monitor.

4.  Refasten the Velcro.

The Leak Detector will periodically (up to 4-minute intervals) draw clean air from an inlet port on the 

side of the chassis for the CO2.
The external Purge/Zero Air Carbon Filter must be connected to the intake port in order to absorb any 

refrigerant present in the area of operation. 

Summary of Contents for PAGM

Page 1: ...User Guide Gas detecting sensor Type PAGM for CO2 and Halocarbon www danfoss com...

Page 2: ...xhaust and zero air filter assembly 3 2 1 Gas Sample Probe 3 2 2 Purge Zero Air Carbon Filter Assembly 3 2 3 Battery 4 3 Front Panel Display Controls 4 4 General operation 5 4 1 Normal operation 5 4 2...

Page 3: ...tions For details on appropriate use refer to the general description application and operation discussions in this manual Before risking equipment damage or personal injury contact the manufacturer i...

Page 4: ...Before the unit can be used please connect the battery pack the unit is designed for a 16 V DC input Set battery voltage output to 16 V refer to the battery manufacturer s instruction sheet for detail...

Page 5: ...ng functions MEASURE indicates that the monitor is actively measuring the refrigerant gas level at the sample point To reset the peak value to zero press the ENTER button PURGE is displayed when the m...

Page 6: ...ng the SILENCE button while a fault condition still exists causes the internal audible alarm to turn OFF for a period of time as set by the SILENCE The front panel FAULT light will continue to flash O...

Page 7: ...act service department for further instructions 0800 Sample Flow Fault Check for a restriction in the gas sample inlet or exhaust Check the external Coalescing Inline Filter and or internal Hydrophobi...

Page 8: ...ied out at least once every 6 months but is dependent on the environment and usage If used in harsh environment where a large amount of dust and moisture particles are present in atmospheres the maint...

Page 9: ...ient CO2 level typically 390 ppm Example unit exposed to 5000 ppm should read 4610 ppm 5 which is from 4380 4840 ppm PAGM IR for halogen gas Set up the unit to measure R134A gas Using a regulator infl...

Page 10: ...Danfoss DCS vt 2021 03 BC372016561395en 000101 10 ADAP KOOL...

Reviews: