Spyder 3 GigE User’s Manual
DALSA 03-032-10158-09
101
Mnemonic
Syntax Parameters Description
set baud rate
sbr
i
Set the speed of camera serial
communication port. Baud rates:
9600
,
19200
,
57600
, and
115200
. Default:
9600.
set ccd direction
scd
i
Sets the CCD shift direction where:
0 =
Forward CCD shift direction.
1 =
Reverse CCD shift direction.
2 =
Externally controlled direction
control via CC3.
set data mode
sdm
i
Sets the camera’s bit width where:
For SG-10-01K40 and SG-10-02K40
0 =
8 bits, 1 tap, 40MHz data rate
1 =
12 bits, 1 tap, 40MHz data rate
For SG-10-01K80 and SG-10-02K80
2
=
8 bits, 2 taps, 80Mhz data rate
3
=
12 bits, 2 taps, 80MHz data rate
set digital offset
sdo
t i
Subtracts the input value from the
video signal prior to FPN correction.
t
= tap selection, either
1
or
2
depending on camera model, or
0
for
all taps.
i
= Offset in a range from
0
to
2048
DN.
set exposure mode
sem
m
Sets the exposure mode:
2 =
Internal SYNC, internal PRIN,
programmable line rate and exposure
time using commands
ssf
and
set
3 =
External SYNC, internal PRIN,
maximum exposure time
4
= Smart EXSYNC
5
= External SYNC and PRIN
6
= External SYNC, internal PRIN,
programmable exposure time
7 =
Internal programmable SYNC,
maximum exposure time. Factory
setting.
8
= Internal SYNC, internal PRIN,
programmable exposure time.
Maximum line rate for exposure time.
set exposure time
set
f
Sets the exposure time. Refer to the
camera help screen (
h
command) for
allowable range.
set fpn coeff
sfc
x i
Set the FPN coefficient.
x
=pixel number within the range
1
to
sensor pixel count
.
i
= FPN value within the range
0
to
2047
(12-bit LSB).
Summary of Contents for Spyder 3 GigE SG-10-01k40-00-R
Page 17: ...Spyder 3 GigE User s Manual DALSA 03 032 10158 09 17 ...
Page 37: ...Spyder 3 GigE User s Manual DALSA 03 032 10158 09 37 ...
Page 43: ...Spyder 3 GigE User s Manual DALSA 03 032 10158 09 43 ...
Page 47: ...Spyder 3 GigE User s Manual DALSA 03 032 10158 09 47 ...
Page 115: ...Spyder 3 GigE User s Manual DALSA 03 032 10158 09 115 ...