Shad-o-Scan User Manual, Rev04
19
Shad-o-Scan 3001/4501 X-Ray Detector
Teledyne DALSA
Setting up the Detector
Network Interface Card
A Network Interface Card (NIC) is required to acquire images and control the Shad-
o-Scan detector. A NIC with Jumbo packet support up to 9014 bytes is required to
receive streaming video, preferably with a PCI Express slot.
Teledyne DALSA advises to use the 5G adapter Network Interface Card (NIC) which
is a low cost but high performance and reliable adapter with a PCI Express slot.
Make sure that the latest driver of the NIC manufacturer is installed. We advise to use
a 64-bit operating systems as the NIC drivers are better maintained.
Due to the high bandwidth used by the detector we strongly advise to have a direct
connection from the NIC to the detector.
Connect Trigger Cable
Connect the trigger cable to the detector when the Trigger Mode of the detector is set
to Synchronous or Triggered.
Connect Ethernet Cable
Connect the CAT-6 Ethernet cable from the detector to the computer Ethernet jack.
Secure the RJ-45 connector to the detector.
Connect Power Cable
Connect a power cable from the detector to a power supply with a supply voltage
between 11-16V and capable of delivery 21W.
Status Light
In normal operation the status light is off.
The status light is red when the detector is in start-up mode.
When the detector is connected to a host system it will be recognized in
approximately 30 to 60 seconds.