background image

 

 

2.  GAS DETECTOR CALIBRATION 

 
 
 
For gas detectors, calibration consists in placing the sensor cell of the detector in an atmosphere made up of air 
and calibration gas.  
 
According to the surrounding atmosphere conditions, applying pure air (zero gas) to the sensor cell may be 
necessary. 
 

2.1.  Toxic gases & oxygen detector (DTX CAN) 

 
 
To perform a correct calibration of the DTX CAN detector you need : 
 

1.  A zero gas cylinder (pure air or Nitrogen); 
 
2.  A calibration gas cylinder (concentration between 1/3 and 2/3 of measuring range); 

 

3.  A pressure reducer with flowmeter; 

 

4.  An injection cap (Figure 4); 

 

5.  A Teflon or polyurethane flexible hose (Ø 7mm); 

 

6.  An insulated flat screwdriver (width 1mm). 

 
 
In order to avoid ambient air or air currents from corrupting the calibration atmosphere around the cell always use 
the suitable injection cap. Just place it instead of the metal detection head cap and fix its position by mean of the 
fastening screw (Figure 4

Figure 4 : Injection cap for DTX CAN

). To get the injection cap please contact Dalemans. 

 
To apply zero gas or calibration gas, fix one end of the flexible hose to the cylinder by mean of the flowmeter. Fix 
the other end to the injection nozzle.  
 
 

Flowmeter 

Hose Ø7mm 

(Teflon or PU) 

Fastening screw 

Injection nozzle 

Injection cap 

Cylinder

 

 

Figure 3 : Zero gas or calibration gas cylinder 

 

 
 
To perform calibration always proceed in the following order : 
 

 

Set the zero of DTX CAN (Zeroing); 

 

 

Adjust sensibility of DTX CAN (Gain). 

Can Bus Gas Detectors - 

Man-CanDetectors-Cal-En-V10.doc

 

C lib

ti

Summary of Contents for octoPRO

Page 1: ...CAN Bus System Calibration manual for Toxic Flammable gases detectors Gas detection w w w d a l e m a n s d e t e c t i o n c o m...

Page 2: ...mans S A can not be liable for direct or indirect damages arising out of the non observance of its instructions Use only Dalemans original parts when you assume the maintenance of the material as desc...

Page 3: ...ON 2 2 1 Toxic gases oxygen detector DTX CAN 2 Zeroing the DTX CAN 3 Adjusting gain of DTX CAN 3 2 2 Flammable gases detector DEX CAN transmitter 4 Set the cell operating voltage 5 Zeroing the DEX CAN...

Page 4: ...19 54 55 34 Email info dalemans detection com In a Dalemans CAN bus system based on OctoBUS 64 central measuring unit to perform the complete calibration procedure of the detectors you need the hand h...

Page 5: ...rewdriver width 1mm In order to avoid ambient air or air currents from corrupting the calibration atmosphere around the cell always use the suitable injection cap Just place it instead of the metal de...

Page 6: ...Instruction Manual for measure function Figure 5 Adjusting GAIN on DTX CAN Adjusting gain of DTX CAN Detector must have been zeroed as described here above On the cylinder set flow rate between 0 4 t...

Page 7: ...ion nozzle The injection nozzle must be fitted in the detection head opening Figure 8 Figure 8 Applying gas to the detector Man CanDetectors Cal En V10 doc Can Bus Gas Detectors C lib ti Figure 6 Cali...

Page 8: ...CAN transmitter Power on the DEX CAN transmitter and wait for at least 15 sec Set the voltmeter on Vdc position and connect it between terminals A and P of the DEX CAN transmitter Note the transmitter...

Page 9: ...ction Manual for measure function Adjusting gain of DEX CAN Detector must have been zeroed as described here above Visualize the detector measure on the OctoPRO display press F2 or on the OctoBUS 64 c...

Page 10: ...NOTES Gas detection...

Page 11: ...il internet R u e J u l e s M l o t t e 2 7 4 3 5 0 R e m i c o u r t B e l g i u m Tel 3 2 0 1 9 5 4 5 2 3 6 Fax 3 2 0 1 9 5 4 5 5 3 4 Gas detection info dalemans detection com www dalemans detection...

Reviews: