MX-5 DL Ultrasonic Thickness Gauge
13
Probe Zero
Setting the Zero Point of the
MX-5 DL
is important for the same reason
that setting the zero on a mechanical micrometer is important. If the tool is
not "zeroed" correctly, all of the measurements the tool makes will be in
error by some fixed number. When the
MX-5 DL
is "zeroed", this fixed
error value is measured and automatically corrected for in all subsequent
measurements. The
MX-5 DL
may be "zeroed" as follows:
Performing a Probe-Zero
1) Make sure the
M X-5 DL
is on.
2) Plug the transducer into the
MX-5 DL
. Make sure that the
connectors are fully engaged. Check that the wearface of the
transducer is clean and free of any debris.
3) The metal probe-disc is on the top end of the
MX-5 DL
. Apply a
single droplet of ultrasonic couplant to the face of this disc.
4) Press the transducer against the probe-disc, making sure that the
transducer sits flat against the surface. The display should show
some thickness value, and the Stability Indicator should have
nearly all its bars illuminated.
5) While the transducer is firmly coupled to the probe-disc, press the
PRB-0
key on the keypad. The
MX-5 DL
will display "Prb0" while it
is calculating its zero point.
6) Remove the transducer from the probe-disc.
Summary of Contents for MX-5
Page 4: ...Dakota Ultrasonics blank page...
Page 6: ...Dakota Ultrasonics 2 blank page...
Page 37: ...MX 5 DL Ultrasonic Thickness Gauge 33 blank page...
Page 39: ...MX 5 DL Ultrasonic Thickness Gauge 35 blank page...
Page 43: ...MX 5 DL Ultrasonic Thickness Gauge 39 blank page...
Page 45: ...MX 5 DL Ultrasonic Thickness Gauge 41 blank page...
Page 48: ......