MX-5 DL Ultrasonic Thickness Gauge
27
Clearing an Entire File
The user may require the contents of an entire file be completely
cleared of all measurements. This would allow the user to start a new
list of measurements starting at storage location L001. The procedure
is outlined in the following steps:
1) Press the
ON/OFF
key to power up the
MX-5 DL
.
2) Press the
MEM
key to activate the data logging functions and
settings.
3) Press the
SEND
key to enter file setup.
4) Press the
UP
/
DOWN
arrow keys to scroll to the file that will be
cleared of all measurements.
5) Press the
SEND
key once again to select the file.
Note:
§
The display will flash the
FILE / F-05
( The file selected by the user
) symbol.
6) Press the
UP
/
DOWN
arrow keys to scroll to the
flashing CLr / F-
05
( The file selected by the user ) symbol.