. Effect using the unit25
4. Effect using the unit
4.1 Intended Purpose
This product including its associated ACCESSORIES intended for the performance of surgical
operation, such as the CUTTING or COAGULATION of biological tissue by means of high
frequency (H.F) current.
4.2 Modes for Monopolar Cutting
1) Pure
- Waveform with minimum thermal damage and hemostasis.
- None modulated 400KHz alternating current.
- For only cutting.
2) Blend 1
- 400KHz alternating current modulated by 33KHz.
- For cutting with minimum hemostasis.
3) Blend 2
- 400KHz alternating current modulated by 33KHz.
- For cutting with more hemostasis than Blend 1.
4) Blend 3
- 400KHz alternating current modulated by 33KHz.
- For cutting with heavy hemostasis than Blend 2.
4.3 Modes for Monopolar Coagulation
1) Forced Coagulation
- 400KHz alternating current modulated by 33KHz.
- For only coagulation.(Standard coagulation)
- Use Ball type electrodes or Mono-polar Forceps.
- Use in state the electrode contacts with tissue.
2) Soft Coagulation
- 400KHz alternating current maximum modulated by 33KHz.
- For only coagulation.
- Minimize tissue adhesion by
‘Time restricted output’.
Summary of Contents for Meditom DT-300S
Page 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches...
Page 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage...
Page 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S...
Page 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage...
Page 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S...
Page 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage...
Page 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S...
Page 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage...
Page 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage...