SVLC-212-175
31
(3) 131H~132L Voltage detection setting
Sets the upper limit alarm value and lower limit alarm value of instant voltage.
131H
Instant measurement voltage
upper limit value
132L
Instant measurement voltage
lower limit value
RESET/SHIFT
MAX/MIN
RESET/SHIFT
MAX/MIN
◆ 131H Instant measurement voltage upper limit value.
132L Instant measurement voltage lower limit value.
Sets the upper limit alarm value and lower limit alarm
value of instant voltage.
The setting range is 30 to 150% (1% step) and OFF.
(To full scale =150%)
Selection by + and - , set value is updated by SET .
Default setting:OFF〔No operation〕(upper limit value,
lower limit value)OFF 80%
(4) 141~142 Backlight setting
Sets the action and brightness of backlight.
141
Backlight action
RESET/
SHIFT
MAX/MIN
RESET/
SHIFT
MAX/MIN
142
Backlight brightness
◆ 141 Backlight action
It can select from ON (always-on), AUTO (auto off), and OFF (always-off)
about action of backlight.
If 5 minutes elapses without operating a switch in case it is set as
"AUTO (auto off)", backlight will go out automatically.
After that, backlight will be turned on if either of switches is
operated.
Selection by + and - , set value is updated by SET .
Default setting:AUTO (Auto off)
◆ 142 Backlight brightness 【white backlight products only】
It can select the brightness of backlight as five steps of 1 to 5.
Backlight becomes the darkest if it is set as "1".
Backlight becomes the brightest if it is set as "5".
Selection by + and - , set value is updated by SET .
Default setting:3 (Middle)
Instant measurement voltage
upper limit value.
Setting factor Now setting
Setting No.
AUTO:
Auto off
+
-
OFF:
Always-
off
+
-
ON:
Always-
on
+
-
Setting
Brightness
5
Bright
4
3
2
1
Dark
Backlight
Setting No
Now setting
New setting
Setting factor