3. Product Usage
3.3 Usage Instructions
-
If turning ON the power of this product, please turn ON the Main Power Switch after turning
ON the Leakage Breaker located at the bottom of the rear panel.
-
If turning OFF the power of this product, please turn OFF the Leakage Breaker after turning
OFF the Main Power Switch. If you turn OFF the Leakage Breaker without turning OFF the
Main Power Switch, the operation of the controller will be continued by the battery built into
this product. This system is prepared for the controller being operated by the battery
built into this product, and the alarm and LCO
2
spraying can be operated when there is a
problem in the power supply such as a power outage.
-
Please start operation after at least one minute when you turn ON the power supply again.
-
The operation of the refrigerator will start operating after 1 minute after power
Since then, the refrigerator is also alternately starts and stops at predetermined intervals.
-
All the operations should be operated through the touch screen.
3.3.1. TEMPERATURE Setting and Operation
-
When the product is “ON”, then the temperature thereof will be automatically rising or
falling toward the set value.
-
By pushing the “SET” button located on the right of the display of set temperature, the
‘Temperature Setup’ window will be popped up wherein users may set desired temperature
with the value inputted directly through the numeric ‘Button’.
<Fig.3.3.1>
Set up of Operating Temperature