![Daihan Scientific WiseSpin CF-10 Operation Manual Download Page 16](http://html1.mh-extra.com/html/daihan-scientific/wisespin-cf-10/wisespin-cf-10_operation-manual_3237137016.webp)
DAIHAN-Sci.com / ALL
for
LAB.com
WiseSpin
®
CF-10 Personal Table-top Centrifuge
16/37
value and TIMER lamp
○
12
is ON. (rpm SV on the upper line of LCD changes to the value of
remaining time). If you set timer SV to “0” before running the rotation, rotation will operate
continuously without stopping unless you stop the rotation operation by clicking jog-shuttle knob
in Set Rotation Speed mode.
If you want to change the rpm SV, you have to stop the centrifugation by clicking jog-shuttle
knob again. (Remaining time display changes to rpm SV. ROTATION lamp
○
11
is blinking and
TIMER lamp
○
12
is OFF). Once stopping, you can adjust rpm again, and then click the jog-shuttle
again to restart. In this case, timer will be reset to its SV, and the timer restarts from the
beginning as well after reaching to target RPM.
(NOTE)
Control resolution is 100 rpm and maximum rpm is 13,500.
Stop and Restart
: If you click the jog-shuttle knob during rotation operation in Set Rotation
Speed mode, the unit stops rotation and it completely stops after approximately 15 seconds
(When you set SV to maximum specification – 13,500 rpm). If you click again in this stop status,
the unit will restart to operate.
Saving current SV
: You could save the current rpm SV for re-use in the next time by pushing
jog-shuttle knob for one second. The unit stores the set value in nonvolatile memory, so you
could re-use it afterwards.
4. Set Timer Mode
You can use internal digital timer to make the operation preserved for a certain period of time. If
you push MODE button
○
13
in Set Rotation Speed mode, you can enter into the Set Timer mode.
You can find zero or non-zero value of the timer SV on LCD and TIMER lamp
○
12
is blinking.
z
If the SV is non-zero, it means that someone had saved a value previously by storage
function. You can use this saved value or you can modify the value by turning the jog-
shuttle knob
○
15
.
z
If the SV is 0, it means there is not value saved previously. You can adjust the timer SV
by turning jog-shuttle knob
○
15
.
Timer adjustment and running Timer
: When you turn the jog-shuttle knob
○
15
for adjustment
Summary of Contents for WiseSpin CF-10
Page 1: ...WiseSpin CF 10 Personal Centrifuge Operation Manual Version 1 4 2 written for Model CF 10...
Page 36: ...DAIHAN Sci com ALLforLAB com WiseSpin CF 10 Personal Table top Centrifuge 36 37...
Page 40: ...DAIHAN Sci com ALLforLAB com WiseSpin CF 10 Personal Table top Centrifuge 40 37...
Page 42: ...DAIHAN Sci com ALLforLAB com WiseSpin CF 10 Personal Table top Centrifuge 42 37 8 1 2 3 4...