DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Ultra Low Temperature Smart Freezer 16/40
2) If press and hold “Set” and Auto/Manu button in “Operation Mode”, it goes into
“Controller Configuration Mode”. The set value in “Controller Configuration Mode” is
extremely important to operate and already set in QC testing process in our factory. It
may cause abnormal operating and technical problem due to the set value has been
changed. Therefore we do not recommend to arbitrarily change the set values in
“controller configuration mode”. If the freezer goes into “controller configuration mode”
by accident, please push and hold “Set” button to go back to operation mode.
If the freezer does not work after changing certain set value in “Controller Configuration
Mode”, please contact the nearest our official distributor or DAIHAN Scientific Co. Ltd.
for A/S request.
1.
LCO
2
Injection
WUF Ultra Low Temperature Freezer uses the high qualified parts and has the simplest structure
for superior temperature stabilization in the chamber. However, in case of sudden power failure,
temperature in the device could be increased and it may cause damage of samples in the
chamber.
For prevention of the abnormal case, we offer LCO2 injection system optionally. When
temperature is abnormally increased, LCO2 gas is automatically dispersed and the system beeps.
ERROR Message
Causation
Solution
Temp Error
When display show unreasonable temp.
value by out of order of temp. sensor
Call the A/S Center.
Temp Sensor
Error
When temp. sensor is on abnormal.
Call the A/S Center.
Door Open
When over 1 min. after open the Door.
Check status of the Door and if
still happen it even door closed,
call the A/S center.
Power Error
When power supply is on abnormal.
Check the Circuit Breaker and
if still happen it even leakage
breaker ON, call the A/S center.
<Table.3.3.1> Error Message Table