DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Forced Convection type Oven
19/48
-
The scope of the compensative value is -20.0
℃
~ +20.0 , and it is possible to adjust at
℃
0.1
℃
unit.
-
After setting the compensation value, the indication of the current temperature is displayed
as the temperature including the compensation value.
-
When entering into the compensation value setting mode during the Heating operation, the
Heating operation that is operating is stopped. After setting the compensation value, the
Heating operation is re-conducted in the Set Heating Mode.
-
If the PUSH button of Jog Dial Knob is pressed and held for 2-3 seconds, the setting value
is saved and this saved value is maintained even when the power is turned on again.
-
If pressing Mode button, it is switched to the Set Heating Mode.
③
MAIN - Safety Locking
-
If pressing MAIN button in the Set Temperature Mode, it is switched to the Locking mode.
-
In the Locking Mode, the Turning by the Jog-Dial Knob or Push button do not applied, and
LCD Display and LED lamp indicate the status that is currently operating.
-
If pressing once the MAIN button, it is entered into the Set Operation Time Mode.
* Automatic Operation Function
In the use of the product, if the power supply has been interrupted and then is reapplied, it has
the function that protects the experimental results and samples by automatically reactivating as
the setting of final operation.
In Jog-Dial operation method, support this function by activating the setting temperature stored
through the saving method of the set points.
When reapplying the power supply according to the problem on safety or user’s intention, if you
want to disable the automatic operation function, set up the temperature to zero (0) according
to the instruction of Jog-Dial.
< Fig 3.3.5. Safety Locking >
Temp PV
: 27
℃
Time PV : 00:00
Present
Temp
Remain Time