Note
:
For the blackbody display of the LED, just press the bottom
‘︿’‘﹀’
to adjust
the value; for the LCD display device, press the bottom
‘<’ to select the value to be
adjusted, and then press the bottom
‘︿’‘﹀’
to adjust the value.
b)
After fixing the camera and the blackbody position, make sure that the blackbody
is on the right or left side of the image and is not blocked (the best position of the
blackbody: put it on the top left or the middle position of the ‘Cross type’ on the
top right: as shown below)
c)
(
In the procedure before 20200210
, the rule window should be in the middle of
the black body, the smaller the window is, the higher the measurement accuracy is,
refer to the green window in the upper right corner of the image below) fill in the
black body temperature 35, double-click the image to enlarge to the full screen
when drawing the area (double-click again after drawing to restore the original
picture size), the rule window should be in the middle of the black body radiation
surface, and the rule window should be as small as possible, and click OK to take
effect;