8. Command mode
Please set up the device No. referring to INDICATOR Manual.
(Can setup from 1 to 32 channel.)
1. Command form (PC -> INDICATOR)
CODE
BYTE1
BYTE2
BYTE3
BYTE4
BYTE5
ASCII
I
D
0
1
P
HEX
49H
44H
30H
31H
50H
1) BYTE
2) BYTE
umb
3) BYTE
1, BYTE2 : Fixed character (ID)
3, BYTE4 : device n
er (1 ~ 32)
5
: command order (P, H, R, Z)
2. Command chart
Command
ASCII
HEX
P
50H
Transmit the current value of order equipment.
H
48H
Hold for order equipment.
R
52H
Release hold for order equipment.
Z
5AH
Operate the current value of order equipment as ZERO.
3. Transmission DATA form (INDICATOR -> PC)
CODE
BYTE1 BYTE2 BYTE3
BYTE4
BYTE5
BYTE6 BYTE7 BYTE8
ASCII
I
D
0
0
1
,
+
0
HEX
53H
54H
30H
30H
31H
2CH
2BH
30H
CODE
BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16
ASCII
1
2
3
4
.
5
CR
LF
HEX
31H
32H
33H
34H
2EH
35H
0DH
0AH
1) BYTE
2) BYTE
3) BYTE
4) BYTE
5) BYTE
CARR
1, BYTE2 : Fixed character (ID)
3 ~ BYTE5 : Device number (1 ~ 32)
6
: Fixed character (,)
7~BYTE14 : DATA 8byte (inc/-)
19
15
:
IAGE RETURN