Other Setting
EMPTY Range setting
F80
X.X.X.X.X.X.
(0.0.0.0.1.0)
You can set “EMPTY” Range.
Within set range, indicator will not display current weight and just
display “Zero”.
“0.000” setting : When Net Zero, “Zero” status lamp and Near Zero
relay will be output.
“0.190” setting : Within 190, “Zero” Status lamp and Near Zero relay
will be output.
Analogue Output Setting (only for the analogue option installation)
F83
XXXXXX
At the set weight value, analogue output will be maximized.
Ex.) Set 5000, then a weight reached 5000
20mA or 10V will be
output
But if you need just 3000 of Max. capa, you can input 3000 through
this function, then the weight reached 3000
20mA or 10V will be
output
Span Value Check
F89
XXXXXX
At this function, you can check the Calculated Span value.
※
If you have difficulty to process Calibration again, the best way to
matching the net weight and display weight is doing Calibration
process once again.
DATE Check / Change
F90
Check Current DATE data or you can Change to new date
TIME Check / Change
F91
Check Current TIME data or you can Change to new TIME
Multi-Calibration Span Value 1 Check
F96
XXXXXX
At this function, you can check the Calculated Span value for multi
calibration 1.
Multi-Calibration Span Value 2 Check
F97
XXXXXX
At this function, you can check the Calculated Span value for multi
calibration 2.
Multi-Calibration Span Value 3 Check
F98
XXXXXX
At this function, you can check the Calculated Span value for multi
calibration 3.