5-3. Simulation Calibration Mode (Without Test Weight) – Mode 2.
- This calibration Method will be useful to make calibration more than 10ton capacity setting.
- Guaranteed resolution will be 1/5,000 and if you need higher resolution, please make
calibration with Test weight.
Step 1. Enter to the “SET-CAL” mode
Press
key
during 5 sec
display.
Or turn on the Power + with pressing
display
Press
key
display.
Press
key to start “Simulation Calibration Mode”.
Remarks : Go to next step with save
key / Back to previous step
key Step 2.
Step 2.
Digit (Division)setting (
display)
Press
key and stop the optimal Digit value.
(Whenever pressing
key, digit value will be changed, like 01-02-05-10-20-50)
Press
key and save change and move to next step.
SET-CAL
TEST
SET-CAL
CAL_ _2
d _ 1
15
/
41