3.5. ADDITIONAL PROCEDURES PRIOR TO EQUIPMENT USE
Even prior to its first use, the equipment must be cleaned and disinfected; the same
additional procedures must be followed for reuse, as described in Chapter 12.
3.6. WARNINGS AND/OR CAUTION TO BE ADOPTED
3.6.1. WARNINGS AND/OR CAUTION DURING EQUIPMENT INSTALLATION
The equipment must be installed only by service technicians authorized by the
manufacturer.
Place the equipment on a site where it will not be in contact with moisture or
water.
Install the unit on a site where it will not be damaged by pressure, temperature,
moisture, direct sunlight, dust or salts.
The equipment must not be submitted to inclination, excessive vibration or shock
(including during transportation and handling).
This equipment has not been designed for use in facilities where vapors,
flammable anesthetic mixtures in contact with air, oxygen or nitrous oxide can be
detected.
Check the equipment’s voltage when performing electric installation.
Failure to
do so may damage the equipment.
The equipment must be properly grounded.
Failure to do so may result in
“Safety Hazard”.
Depending on local regulation the X-Ray emission control may require
installation outside the facility where the equipment is placed, and the operator
may need visual contact with the patient through a window with radiological glass
or similar, since the operator must not lose visual contact with the patient.
Mobile and portable RF communication equipment can affect the Eagle 3D
Computerized Tomography Machine.
This equipment must be solely used by health care professionals as it may cause
radio interference or interrupt the operation of nearby equipment. Mitigatory
measures, such as equipment re-orientation or replacement and the facility’s
screening, may be necessary.
3.6.2.WARNING AND/OR CAUTION DURING EQUIPMENT USE
The equipment must be operated only by qualified and trained professionals
(dentists, radiology technicians, hygienists or engineers).
Always observe the display messages, the equipment as a whole and the
patient in order to detect any arising problems early.
In case occasional maintenance is required, use only services provided by
Authorized Service Technicians.
The equipment has been designed to withstand continual and intermittent
operation; therefore, follow the cycles described in these operation
instructions.
Summary of Contents for EAGLE 3D
Page 1: ...US Mod SER del Eag S MA gle 3D ANUA AL...
Page 2: ...1 THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Page 93: ...92...
Page 95: ...94...
Page 96: ...95...
Page 97: ...96...
Page 100: ...17 7 5 IRRADI IATED FIE ELD SIZE 99 TOMOGR RAPHIC E EXAM...
Page 105: ...18 18 1 IDENT PACKA TIFICAT AGE TION LA 104 ABELS...
Page 106: ...18 2 PRODU UCT 105...
Page 107: ...19 Pow EQUIP wer supply u PMENT unit slim DIMENS 106 SIONS...