17.8. ELECTROMAGNETIC
EMISSIONS
Manufacturer’s guidelines and declaration - electromagnetic emissions
The EAGLE 3D Computerized Tomography Machine is intended for use in the electromagnetic
environment specified below. Should the buyer or user of the EAGLE 3D Computerized
Tomography Machine ensure that it is used in such an environment.
Emission assays
Compliance
Electromagnetic Environment
Guidelines
RF emissions
CISPR 11
Group 1
The EAGLE 3D Computerized
Tomography uses RF energy only
for its internal functions. However,
RF emissions are very low and are
not likely to cause any interference
with electronic equipment nearby.
RF emissions
CISPR 11
Class B
The EAGLE 3D Computerized
Tomography Machine is suitable for
use in all types of facilities, including
residential facilities and those
directly connected to the public
system of low-voltage electric power
supply for residential buildings.
Harmonic emissions
IEC 61000-3-2
Class A
Emissions due to
voltage/scintillation
fluctuation
IEC 61000-3-3
In compliance
17.9. ELECTROMAGNETIC
IMMUNITY
Manufacturer’s guidelines and declaration - electromagnetic immunity
The EAGLE 3D Computerized Tomography Machine has been designed for use in
electromagnetic environments, according to the specifications below. The client or X-Ray
Machine operator must ensure that the equipment is used in such type of environment.
Immunity Assays
ABNT NBR IEC 60601
Assay Level
Compliance Level
Electromagnetic
environment -Guidelines
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV by contact
± 8 kV by the air
± 6 kV by contact
± 8 kV by the air
Floors must be finished
with wood, concrete or
ceramics. In case the
floor is covered with
synthetic material, relative
humidity must be at least
30%.
Fast electric
transients /pulse
train (“Burst”)
IEC 61000-4-4
± 2 kV on the mains
supply line
± 1 kV on the
input/output line
± 2 kV on the mains
supply line
± 1 kV on the
input/output line
The quality of power
supply must be that of
hospital facilities or of
typical business facilities.
Impulses
IEC 61000-4-5
±1 kV - differential
mode
± 2 kV - regular mode
±1 kV - differential
mode
± 2 kV regular mode
The quality of power
supply must be that of
hospital facilities or of
typical business facilities.
Voltage drops,
short interruptions
and voltage
< 5% Ut
(>95% of voltage drop
in Ut) per 0.5 cycle
< 5% Ut
(>95% of voltage
drop in Ut) per 0.5
The quality of power
supply must be that of
hospital facilities or of
Summary of Contents for EAGLE 3D
Page 1: ...US Mod SER del Eag S MA gle 3D ANUA AL...
Page 2: ...1 THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Page 93: ...92...
Page 95: ...94...
Page 96: ...95...
Page 97: ...96...
Page 100: ...17 7 5 IRRADI IATED FIE ELD SIZE 99 TOMOGR RAPHIC E EXAM...
Page 105: ...18 18 1 IDENT PACKA TIFICAT AGE TION LA 104 ABELS...
Page 106: ...18 2 PRODU UCT 105...
Page 107: ...19 Pow EQUIP wer supply u PMENT unit slim DIMENS 106 SIONS...