17
1) Connect the energy plug to the socket.
2) Raise the door latch, push it to the left, and check
the cleanliness of the autoclave.
3) Using the dosing cup (21), pourinside the autoclave
the following quantities of water:
Autoclave partially full - 150ml.
Autoclave completely full - 200ml.
Use only distilled water or demineralized. If you use
other water than distilled, it may cause the
obstruction of the internal valves and stain the
instruments, and the warranty will not be applicable.
4) All material being sterilized must be previously
washed, rinsed and dried.
5) Close the autoclave’s door pushing it against
the tank, pushing the door latch to the right and
then lowering it. Check that it is correctly closed
(23). In order to avoid accidents and burns, it’s very
important that the door is well closed.
6) Green led (4D) in the front panel will come on,
indicating that the autoclave is on.
WARNING:
Don’t use “Kraft / TNT”
paper for wrapping the materials to besterilized
(types of wrappings for autoclave sterilization, see
page 25).
Observe each country’s standards for preparation of
material to be sterilized.
Distribute the wrappings at equal distances (22)
leaving enough space between them so as to allow
the steam to circulate.
ATTENTION:
The material to be sterilized (22) can’t be in contact
with any of the internal walls of the chamber.
OPERATION OF EQUIPMENT