Completed changes should be confirmed by „
tu
” (Factory settings will
be recorded in the meter memory).
Depressing „
u
” we can go to the next menu position.
Depressing „
p
” we leave „
RSCT
”, depressing „
p
” again we come
out of the programming mode and we enter the temperature
measurement mode.
5.3.
Recording results
Determining whether measurement results should be recorded in the
meter memory.
Using „
u
” select „
TMEM
” position in the programming „
MEMO
”
mode.
The auxiliary display indicates the current value of the parameter. To
change it depress „
tu
” (the digit on the auxiliary display will start
blinking).
Using „
p
” set the required value.
The parameter can assume the following values:
!
0
results are not recorded,
!
1
storing in memory consecutive 300 measurements with frequency
determined by „
RSCT”
(see item
5.1
),
!
2
storing in memory consecutive measurements. After memory
overflow measurements are shifted “upwards” (the first position is
deleted, and the (latest) result is recorded in the last position.
ATTENTION
: When the parameter value is
1
the first position of the
auxiliary display presents symbol
Ï
. It means that recording results in
memory is in progress.
The symbol comes off after recording 300 (results of) measurements.
Completed changes should be confirmed by „
tu
” („
TMEM
”
message would start blinking again). Depressing „
p
” we come out of
„
TMEM
”, depressing „
p
” again we leave the programming mode and
enter the temperature measurement mode.