MAINTENANCE 

 
Daily: Wipe off the unit. 
 
Once every two weeks: 
Clean the BATH 
 

1.

 

Remove  the  RACK  and  LINK  and  take  them  to  the  sink,  clean  under  running  water  and  wipe  with 
bleach towelettes or a 10% bleach solution. Rinse off in running water. 

2.

 

Drain the bath by attaching the provided tubing to the DRAIN valve. [ See page 9 ] Tilt the bath to drain 
as much water as possible through the drain valve.  

3.

 

Wipe the BATH with bleach towelettes or a 10% bleach solution. Use only water and bleach towelettes or 
a 10% bleach solution for cleaning. Use alcohol for removing oil or grease.  DO NOT IMMERSE THE 
WHOLE BATH.  WIPE THE PUMP SCREEN FROM THE OUTSIDE ONLY. DO NOT REMOVE 
THE SCREWS. Rinse out the bleach with water. 

4.

 

Remove  the  drain  tubing  and  fitting  to  close  the  DRAIN  valve.    Fill  the  bath  with  water.  If  you  use 
deionized  or  distilled  water  add  a  little  pinch  (about  0.01  gr.)  of  table  salt.  .      Add  Algaecide  Part 
#ALG240 (1.0 ml. per 4 liter of water. Available USA only) to the water. Any other bath clearing agent 
can also be used. 

5.

 

Mount the RACK 

 
 
Every 3 Months: 
Temperature Calibration. Run temperature calibration at least every 3 months or more often if your institution 
procedures or regulations require. See PROGRAM MODE > TEMPERATURE CALIBRATION. 
 

Overtemperature Alarm Test.  Hold MINUS as you turn the unit ON.  DIAGNOSTICS will be displayed. Push 
ENTER. Push and release PLUS until  ALARM TEST is displayed.  Push ENTER, the heater will heat 
continuously and the pumps will turn on. When the temperature reaches 1.0C above the target temperature the 
alarm will sound. Turn Off the unit.  If alarm does not sound, contact CytoTherm. Alarm test should be 
performed every 3 months as part of unit maintenance.

 

 

PARTS LIST  CT-4S 

 

14-MP 

Magnetic pump 

DP-SW   

Switch, 2-pole    

14-77 

Heater (2 needed) 

C-TH 

Safety Thermostat 

CT4-PC 

Printed Circuit Board 

C4-DISP 

Display 

C4S-BTH 

Bath Section 

C4-TR 

Transformer 120V 

C4S-RKM 

Rack Motor 

C4-TR2 

Transformer 240V 

C4-TS 

Temperature Sensor 

C4S-CLM 

Clamp  ( bag to rack) 

C4S-EL 

Evaporation Lid 

C4S-RK 

Rack (holds 4 bags) 

C4S-LINK 

Link Bar 

C4S-PM 

Program Module 

C-VLVF 

Drain Valve  Female 

ALG240 

Algaecide. USA only 

C-VLVM 

Drain Valve  Tubing 

 

 

Always specify model and serial number when ordering.

 

 

SOURCES for SUPPLIES 

BAGS overwrap 15 cm. wide by 30 cm. (6” x 12”) long are available from US Plastics 1390 Neubrecht Rd. Lima OH 45801 
Tel. 800 537-9724. Part #47352. 

Summary of Contents for CT-4S

Page 1: ...ax 609 396 9395 Please mail in your warranty card Please record the following Serial No ___________ Date of Purchase____________ TABLE OF CONTENTS Draw Instructions CT Products CT4SinsM doc 26 Jun 16...

Page 2: ...nit comes packed in 1 box Make sure you have received the following Water Bath Link bar Rack with 4 bag clamps Drain Tubing Evaporation Lid Power Cord Instruction Book Overwrap bags Algaecide USA only...

Page 3: ...products thawed without an overwrap bag Hold a bag frozen plasma bare or overwrapped with the tubes up Slip the CLAMP plastic piece with a stainless spring attached over the top of the bag and line up...

Page 4: ...d tubing to the DRAIN valve See page 9 Tilt the bath to drain as much water as possible through the drain valve 3 Wipe the BATH with bleach towelettes or a 10 bleach solution Use only water and bleach...

Page 5: ...t maintenance PROGRAM MODE Allows you to 1 Program the thawing time 2 Calibrate the temperature 3 Set the target thawing temperature The Alarm temperature is 1 0C above the target temperature 4 Select...

Page 6: ...H and the heaters 14 77 for continuity Temperature calibration does not hold it changes Replace temperature sensor CT4 TS Temperature calibration is out of range Have technician check the voltage at T...

Page 7: ...erature calibration at least every 3 months or more often if your institution procedures or regulations require See PROGRAM MODE TEMPERATURE CALIBRATION Overtemperature Alarm Test Hold MINUS as you tu...

Page 8: ...er OVER TEMPERATURE The unit will display TOO HOT and emit double beeps if the temperature reaches 1 0o C above the thawing temperature The controls will turn the heaters off In the unlikely situation...

Page 9: ...gently pinch the drain tube and release to remove any air to improve flow 9 CT 4S Quick Connect Drain Insert drain tube into drain housing part C VLMF until it engages Push in and turn clock wise to l...

Page 10: ...2Blu Vio 2Grn 2Yel 2Orn 2Red 2Brn Ribbon Cable CT 4D Board Blk Blk Blu Blu B r n V i o 47 uF 250 V Blk Blk Whi Whi Whi Red May 2002 Rack Motor Blu Blu Both Blu wires are Neutral Used in CT 4S Instruc...

Reviews: