MODES 

 

STANDBY MODE.  Just turn the unit ON. The unit recirculates the water in the BATH, The unit heats up and 
maintains 37.0

O

C in the BATH.   

 

THAWING MODE. The unit continues recirculating tempered water through the BATH and  starts moving the RACK 
back and forth. The hanging plasma bags are pushed against rods in the RACK mixing the plasma in the bags.  A count-
down timer is started to display the time-to-completion of thawing. START initiates thawing. Timing out or pushing STOP 
stops thawing. 
 

DIAGNOSTIC MODE. Is for technician access. Hold MINUS as you turn the unit ON.  

1.

 

To restore the unit to initial factory settings, press ENTER  while still holding MINUS. The unit displays 
“UNIT RESET”. Turn the unit off. You have to re-calibrate the temperature after reseting the unit. 

2.

 

“PUMP TEST” is a diagnostic tool to check if each actuator turns on. It is checked only when a pump does 
not operate. Release MINUS then push ENTER to verify that you are a technician. The display shows 
“PUMP TEST”. MINUS turns on Pump 1, ENTER turns the cam to push the RACK back and forth. Push 
PLUS to go to next function. 

3.

 

Overtemperature Alarm Test.  Hold MINUS as you turn the unit ON.  DIAGNOSTICS will be displayed. 
Push ENTER. Push and release PLUS until  ALARM TEST is displayed.  Push ENTER, the heater will heat 
continuously and the pumps will turn on. When the temperature reaches 1.0C above the target temperature 
the alarm will sound. Turn Off the unit.  If alarm does not sound, contact CytoTherm. Alarm test should be 
performed every 3 months as part of unit maintenance. 

 
PROGRAM MODE.  Allows you to: 

1.

 

Program the thawing time. 

2.

 

Calibrate the temperature. 

3.

 

Set the target thawing temperature. The Alarm temperature is 1.0C above the target temperature. 

4.

 

Select the signal when thawing is done (continuous or pulsed). 

 
Enter PROGRAM MODE by holding ENTER while turning the unit ON. Follow the displayed instructions to 
program. Pushing PLUS advances the menu to the next parameter.  
 
Programming the Thawing Time. When the thawing time is displayed, press MINUS because you want to change 
the time. Use PLUS or MINUS to set the desired new thawing time, press ENTER to program in the new time. 
Turn the unit off.  

 

Calibrating Temperature. Turn the unit on and let it heat up to 37.0

o

C.  Check the displayed temperature with 

your accurate thermometer.  Hold or hang your thermometer so the end is in the water in the bath. If the 
displayed temperature needs to be changed, turn the unit off.  Enter PROGRAMMING MODE by holding 
ENTER as you turn the unit on. When the thawing time is displayed push PLUS  until the display shows 
“CALIBRATE TEMPERATURE”. Push MINUS because you want to change the displayed temperature. Use 
PLUS or MINUS to make the displayed temperature agree with your thermometer. Push ENTER to record the 
calibration. Turn the unit off.  If the unit can not be calibrated, or if  the calibrated temperature is not stable,  
have a technician check the unit. See TECHNICIAN FUNCTIONS. 
 

SWITCH  FUNCTIONS    There are 6 switches on the front panel: 

START.  Starts function. 
 
STOP.  Stops function. 
 
ENTER.  Records parameter changes. 
 
PLUS.  Steps through parameters and increases value of parameter. 
 
MINUS.  Allows change in parameter and decreases value of parameter.  
 
Sketch of FIN logo.  Not currently used.  Future expansion. 

Summary of Contents for CT-4S

Page 1: ...ax 609 396 9395 Please mail in your warranty card Please record the following Serial No ___________ Date of Purchase____________ TABLE OF CONTENTS Draw Instructions CT Products CT4SinsM doc 26 Jun 16...

Page 2: ...nit comes packed in 1 box Make sure you have received the following Water Bath Link bar Rack with 4 bag clamps Drain Tubing Evaporation Lid Power Cord Instruction Book Overwrap bags Algaecide USA only...

Page 3: ...products thawed without an overwrap bag Hold a bag frozen plasma bare or overwrapped with the tubes up Slip the CLAMP plastic piece with a stainless spring attached over the top of the bag and line up...

Page 4: ...d tubing to the DRAIN valve See page 9 Tilt the bath to drain as much water as possible through the drain valve 3 Wipe the BATH with bleach towelettes or a 10 bleach solution Use only water and bleach...

Page 5: ...t maintenance PROGRAM MODE Allows you to 1 Program the thawing time 2 Calibrate the temperature 3 Set the target thawing temperature The Alarm temperature is 1 0C above the target temperature 4 Select...

Page 6: ...H and the heaters 14 77 for continuity Temperature calibration does not hold it changes Replace temperature sensor CT4 TS Temperature calibration is out of range Have technician check the voltage at T...

Page 7: ...erature calibration at least every 3 months or more often if your institution procedures or regulations require See PROGRAM MODE TEMPERATURE CALIBRATION Overtemperature Alarm Test Hold MINUS as you tu...

Page 8: ...er OVER TEMPERATURE The unit will display TOO HOT and emit double beeps if the temperature reaches 1 0o C above the thawing temperature The controls will turn the heaters off In the unlikely situation...

Page 9: ...gently pinch the drain tube and release to remove any air to improve flow 9 CT 4S Quick Connect Drain Insert drain tube into drain housing part C VLMF until it engages Push in and turn clock wise to l...

Page 10: ...2Blu Vio 2Grn 2Yel 2Orn 2Red 2Brn Ribbon Cable CT 4D Board Blk Blk Blu Blu B r n V i o 47 uF 250 V Blk Blk Whi Whi Whi Red May 2002 Rack Motor Blu Blu Both Blu wires are Neutral Used in CT 4S Instruc...

Reviews: