CytoTherm CT-4S Owner'S Manual Download Page 1

CytoTherm 

110 Sewell Ave.     Trenton NJ 08610  USA 

Tel 800 747-9699 or 609 396-1456   Fax 609 396-9395 

Email [email protected]    Web  www.cytotherm.com

 

 

 

CytoTherm-4S      

   

Owners Manual 

 

      COMPACT 
PLASMA THAWING 
    WATER BATH 
with Sliding Agitation

 

 

    Model    

CT-4S

 

Version 4.09 and 
higher. 
 

          

 

Warranty 

 
 
Each CytoTherm product is produced under rigid 
quality control standards.  This unit is fully 
warranted for a period of two years from the date of 
purchase.  Call (609) 396-1456 or (800) 747-9699 
for help.  If necessary send unit to: 
 
CytoTherm    110 Sewell Avenue   Trenton, NJ 08610   Fax (609) 396-9395

 

 
Please mail in your warranty card. 
Please record the following:   Serial No. ___________   Date of Purchase____________ 
 

TABLE OF CONTENTS                            

Draw\Instructions CT Products\ CT4SinsM.doc 26 Jun 16

 

  1 

Specifications 

 6 

Troubleshooting 

  2 

Description 

 7 

Maintenance 

 2 

Setting-up 

 7 

Parts List 

 3 

Operation 

 8 

Technician Functions 

 4 

Leaky Plasma Bag 

 8  

Safety Features 

 4 

Cleaning the Bath 

 9 

Draining Instructions 

 5 

Modes of operation,  Programming 

 10 

Wiring Diagram 

 

SPECIFICATIONS:  

Thawing temperature adjustable 25

o

C to 40

C. Preset to 37.0

o

C.  Overtemp alarm 1.0

o

over thawing temperature.  45

o

C safety thermostat.  Accuracy 0.1

o

C. Easy to do, digital temperature calibration. Power 770 

Watt, Available in 120 Volt or 230 Volts, 50 or 60 Hz. (must be grounded). 

 

Summary of Contents for CT-4S

Page 1: ...ax 609 396 9395 Please mail in your warranty card Please record the following Serial No ___________ Date of Purchase____________ TABLE OF CONTENTS Draw Instructions CT Products CT4SinsM doc 26 Jun 16...

Page 2: ...nit comes packed in 1 box Make sure you have received the following Water Bath Link bar Rack with 4 bag clamps Drain Tubing Evaporation Lid Power Cord Instruction Book Overwrap bags Algaecide USA only...

Page 3: ...products thawed without an overwrap bag Hold a bag frozen plasma bare or overwrapped with the tubes up Slip the CLAMP plastic piece with a stainless spring attached over the top of the bag and line up...

Page 4: ...d tubing to the DRAIN valve See page 9 Tilt the bath to drain as much water as possible through the drain valve 3 Wipe the BATH with bleach towelettes or a 10 bleach solution Use only water and bleach...

Page 5: ...t maintenance PROGRAM MODE Allows you to 1 Program the thawing time 2 Calibrate the temperature 3 Set the target thawing temperature The Alarm temperature is 1 0C above the target temperature 4 Select...

Page 6: ...H and the heaters 14 77 for continuity Temperature calibration does not hold it changes Replace temperature sensor CT4 TS Temperature calibration is out of range Have technician check the voltage at T...

Page 7: ...erature calibration at least every 3 months or more often if your institution procedures or regulations require See PROGRAM MODE TEMPERATURE CALIBRATION Overtemperature Alarm Test Hold MINUS as you tu...

Page 8: ...er OVER TEMPERATURE The unit will display TOO HOT and emit double beeps if the temperature reaches 1 0o C above the thawing temperature The controls will turn the heaters off In the unlikely situation...

Page 9: ...gently pinch the drain tube and release to remove any air to improve flow 9 CT 4S Quick Connect Drain Insert drain tube into drain housing part C VLMF until it engages Push in and turn clock wise to l...

Page 10: ...2Blu Vio 2Grn 2Yel 2Orn 2Red 2Brn Ribbon Cable CT 4D Board Blk Blk Blu Blu B r n V i o 47 uF 250 V Blk Blk Whi Whi Whi Red May 2002 Rack Motor Blu Blu Both Blu wires are Neutral Used in CT 4S Instruc...