5.4
Changing tubing holder
Introduction
This section describes how to change the tube holder to connect a tubing of a different
diameter.
Changing tubing holders
The tubing holders can be quickly changed to connect different diameter tubing. The
tubing holders must be changed as a pair (left and right tubing holders must be of the
same size).
The tubing holders are coded so that the instrument automatically reads the tubing
holder type and selects the appropriate program.
Follow the steps below to remove and install the new holders.
Step
Action
1
Open the instrument's cover.
2
Open the existing holder and locate the captive screw.
3
Loosen the captive screw using the M4 Flat screwdriver and pull the holder
off the locating pins.
4
Repeat step 2 on the second holder.
5
To install the new holder, set each new holder on its platform. Align the new
holders to the correct position using locating pins.
6
Tighten the captive screws.
7
Close the holder.
5 Operation
5.4 Changing tubing holder
Sterile Tube Fuser - Wet Operating Instructions 29024010 AD