Step
Action
5
Optional:
Change tubing type.
If you want to change tubing type:
•
Press the
CHANGE button.
•
Press the desired tubing type on the display. If no tubing types appear on
the first screen, press the
NEXT button to open the second screen.
Note:
If you have already inserted the blade and wish to change the tubing type,
simply remove the blade to return to the
INSERT BLADE screen.
6
Optional:
Change tubing dimension.
Use a tubing holder with the desired dimension. The tubing dimension is
then automatically entered in the instrument.
To change a tubing holder, see
Changing tubing holders, on page 51
Insert a cutting blade
Follow the instructions below to insert a cutting blade.
CAUTION
The cutting blade of Sterile Tube Fuser - Wet is a thin piece of stain-
less steel. It does not have a sharpened edge but can cause cuts if
handled improperly. Be sure to read and understand the procedure
for installing and removing the blade.
Step
Action
1
Place the Sterile Tube Fuser - Wet blade in the Installation/Removal Tool as
shown below.
5 Operation
5.3 Fusing tubes
Sterile Tube Fuser - Wet Operating Instructions 29024010 AD