Cygnus MINI ROV Operation Manual
M4-CYGMINI-TSR-M-ENG_Iss 1
Page 54 of 66
Status-byte Structure
Bit Name
Details
7
Gauge Type
=1 (always)
6
Resolution
1 = High Resolution
0 = Low resolution
5
Units
0 = Metric mm
1 = Imperial Inch
4
Reading Range
1 = High Range
0 = Low Range
3, 2 Echo Count
Number of echoes found (0..3)
1
Calibration
1 = Remote (always)
0
No-Reading
1 = No Thickness Reading
0 = Valid Thickness Reading
Data-bytes Arrangement
The four data bytes are ASCII-encoded, decimal thickness value.
Data byte 0 is most-significant. Leading-zeroes are replaced by
ASCII Space [hex 20]. Decimal points are implied from the Status-
byte – Units, Resolution and Range bits:
Reading-type
Hi-Range
Lo-Range
Imperial, Lo-Resolution
x x . x x
x . x x x
Metric, Lo-Resolution
x x x . x
x x x . x
Imperial, Hi-Resolution
x x . x x
x . x x x
Metric, Hi-Resolution
x x x . x
x x . x x