EN-52
Technical Information
Additional equipment connected to medical electrical equipment must comply with the re-
spective IEC or ISO standards (e.g. IEC 60950 for data-processing equipment). Furthermore,
all configurations shall comply with the normative requirements for medical systems (see IEC
60601-1-1 or Section 16 of the 3rd edition of IEC 60601-1, respectively).Anybody connect-
ing additional equipment to medical electrical equipment is a system configurer and is there-
fore responsible for ensuring that the system complies with the normative requirements for
systems. Please note that local laws take priority over the above-mentioned normative require-
ments. If in doubt, consult your local specialised dealer or the Technical Service department.
Storage and transport
conditions between
uses:
If you are storing and or traveling with the device in the recommended zip
bag, the standard condition recommendations apply.
The ambient condition recommendation during use apply to storage and
transport without the zip bag.
Ambient conditions
during use:
Ambient temperature 10 to 35 °C, relative air humidity from 15 to 85 %
(non-condensing), air pressure from 700 hPa to 1060 hPa.
Serial number:
The unit‘s serial number is located inside the battery compartment and
is identified with SN. Please quote this number in the event of possible
service enquiries.
Battery:
Lithium polymer battery, 3.7 V, 1600 mAh, 5.92 Wh. The battery may only
be charged with the supplied charger model SW06-050E.
Calibration:
The unit does not have to be calibrated.
Protection class IP:
IP 22: Protection against solid foreign bodies with a diameter of more than
12.5 mm; protection against small amounts of liquid.
Minimum expected
service life:
Unit: 5 years, power pack: 5 years, sensor: 3 years, battery: 500 charge/
discharge cycles