Clamp:
Please keep the clamp clean. After long time use, please use soft cloth together
with alcohol to erase the dirt. Please keep the clamp away from damp environment, not
to mention direction contact with water.
Handle:
Please use clean cloth or sponge to clean the handle. Please use little alcohol
instead of water to erase the dirt and parch it.
To maintain the performance of the product, there can be one check or calibration every
year.
8. Troubleshoot
Problem
Reason
Treatment
No output signal or
amplitude is low
Power is off
Turn on the power.
Oscilloscope is AC mode
Please set it to DC
mode
Clamp isn’t closed
Check the clamp, close
it completely.
Powerindicator is off
when the probe is on.
Battery voltage is lower
than 6.5V
Change the battery
Amplitude is half
The input impedance is 50Ω
Set it to 1MΩ
9. Packing List
Packaging
Name
Quantity
Current Probe
1
9V Battery
1
DC12V/1.2A Adaptor
(
CK-612)
1
BNC Output Cable(CK-310)
1
Tool box
1
Instruction book
1
Warranty Card
1