TENSPros
13.2
Declaration -
electromagnetic
immunity
Guidance and manufacturer's declaration-electromagnetic
immunity
The SoundCare™ Plus device is intended for use in the electromagnetic
environment specified below. The customer or the user of the
SoundCare™ Plus should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD)
lEC 61000-
4-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30
%.
13.3
Declaration-
electromagnetic
immunity
Guidance and- manufacturer's declaration. Electromagnetic
immunity
The SoundCare™ Plus device is intended for use in. the electromagnetic
environment specified below. The customer or the user of the
SoundCare™ Plus should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted
RF lEC
61000-4-6
Radiated RF
lEC 61000-
4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
3 Vrm
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the
SoundCare™ Plus device,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
separation distance.
d
=1.2 P
d
=1.2 P
d
=2.3 P
80MHz to 800MHz
80MHz to 2.5MHz
19
SoundCare™ Plus Pain Relief Device