TENSPros
Caution
The massage performed with the head should be applied with a regular
movement, not too slow to avoid inducing heat, nor too fast to prevent a
bad contact which would reduce the effectiveness of the treatment.
Warning
If you want to replace treatment head, you need to turn the power switch
OFF, so that the device is in shutdown state in order to the replacement of
treatment head.
6.1.3
After the
treatment
Clean the skin of the area treated as well as the treatment-head by
using a towel or a tissue.
The treatment-head should be cleaned up with a 70% alcohol solution
Check if there are any signs of improvement (e.g. pain, circulation and
mobility).
When entering the next treatment-session, the patient is instructed to
report any possible reaction.
6.2
Operating
the
apparatus
6.2.1
Switch on the
apparatus
Turn the apparatus on with switch. The apparatus executes a self-test;
checking all important functions and presents itself with the start up
settings. The start up settings is adjustable at memory location. To
personalize the start up settings, refer to chapter 5.5,
Therapy system
set up.
6.2.2
Adjusting
parameters
Press the buttons [1], [2], [3], [4] to select a parameter.
13
SoundCare™ Plus Pain Relief Device
6.1.2
During
the
treatment
The treatment-head has to be moved constantly in a slow tempo, also
in case of the semi-statically method.
Ask the patient regularly about his/her experiences. If necessary, the
treatment can be adjusted. By using knob [6], the intensity can be
reduced.
In case of indications of pour transmission of ultrasound energy, it is
advised to add the contact-gel or reposition the ultrasound-head.
During the treatment, if the ultrasonic applicator connected well, the
applicator LED will light; if the ultrasonic applicator doesn't contact
well, the applicator LED will blink slowly,the system will be think this is
no-load in the applicator. When treatment was pause, the applicator
LED light will be extinguished. And the countdown will also be
suspended.