25
4.Some programs will seem
intermittent. This is expected.
Stimulation
is ineffective.
Improper electrode
and applicator
placement
Unknown
Reposition electrode
and applicator
Contact clinician.
The skin
becomes
red and/or you
feel a stabbing
pain
Use the electrodes
on the same site
every time.
Re-position the electrodes. If
at any time you feel pain or
discomfort stop use
immediately.
The electrodes
aren't stuck onto
the skin properly.
Ensure the electrode is stuck
securely on the skin.
The electrodes are
dirty.
Clean the electrode pads with
a damp, lint free cloth or
replace new electrode pads.
The surface of the
electrode was
scratched.
Replace new electrode.
The electrode pads
come off the skin.
Turn off the device and stick
the electrode pad firmly to
the skin.
The cable is
disconnected
Turn off the device and
connect the cable
The power of the
batteries has been
exhausted.
Please replace them with new
batteries.
Output current
stops during
therapy
Summary of Contents for InTENSity 5000 Hybrid
Page 1: ...INSTRUCTION MANUAL ...