Curbell CSM-BC500 Quick Setup Manual Download Page 2

TEST THE ALARM:

1. When pressure is placed back on the pad, the green LED will blink two times 

and then continuously every three seconds indicating the pad is in use.

2. Remove pressure from the pad to activate the alarm. The red LED will then 

blink (for 10 seconds) and nurse call will alarm.

3. To cancel the alarm, press the monitor’s RESET button and cancel the nurse 

call alarm at the patient station.

Insert the nurse call plug directly 

into the ¼" patient station nurse 

call receptacle.

OR

Securely mount to the side of the patient 

station and connect it with a jumper 

cable (sold separately).

Position the sensor pad on 

the chair or bed as instructed/

illustrated on the pad.
Test the alarm.

OPTIONAL — USE AN AC ADAPTOR:

Connect an AC Adaptor (sold separately) to the 

AC input on bottom of the monitor

NOTE: The installation of batteries is required 

even if using an AC Adaptor. The monitor will 

not function without power.

CORDED PAD CONNECTION:

1. Plug the sensor cord into the monitor’s sensor jack.

2. Both LEDs flash 3 times simultaneously. 

Note: If synchronized to cordless pad(s), they will remain synchronized and 

able to be placed in service in tandem with the corded pad.

Check the date on the sensor pad 

intended for use and verify that is 

not beyond the expiration date.

The RESET button provides the following two functions: it cancels the alarm 

or places the monitor into a pause mode that allows the caregiver to take the 

patient out of a bed or chair without signaling the alarm.

1. While patient is on the pad, press the RESET button on the top of the 

monitor to enter the 30 second pause mode. 

2. The yellow and green LEDs will flash simultaneously, and the patient may 

be removed from pad within 30 seconds without activating an alarm. The 

monitor will then stay in the pause (standby) mode until pad is back in use 

(pressure applied) and the monitor system will reactivate.

OPTIONAL — INSERT THE CALL CORD:

1. Insert nurse call cord (sold separately) into 

nurse call cord ¼” jack.

2. Turn the CALL CORD switch to ON

NOTE: Whenever removing call cord from 

monitor, turn the call cord

switch to off.

1

1

1

1

2

2

3

4

2

CORDED SENSOR USE

SETTING UP THE SYSTEM

THE PAUSE FEATURE

CONNECTING THE NURSE CALL CORD AND AC ADAPTOR (OPTIONAL)

1

2

3

1

2