Curbell CSM-BC500 Quick Setup Manual Download Page 1

Open the battery compartment:
1. Install three “AAA” Batteries
2. Close cover and secure screw
Optional: connect an AC adaptor to 

extend battery life.

TEST THE ALARM:

1. When pressure is placed back on the pad, the green LED will blink two times and then continuously every three seconds indicating the pad is in use.

2. Remove pressure from the pad to activate the alarm. The red LED will then blink (for 10 seconds) and nurse call will alarm.

3. To cancel the alarm, press the monitor’s RESET button and cancel the nurse call alarm at the patient station.

Turn the CALL CORD switch to OFF before initially 

inserting the nurse call plug or cable into the patient 

station. If using with the call cord, switch to ON after 

connected to the patient station and inserting the 

call cord. 

Remove pressure from pad(s). 
NOTE: If a pad is in use (pressure applied), it will remain synchronized and 

monitoring the patient, even after attempting to unsynchronize.

Press and hold the black programming button located on the bottom of the monitor 

for approximately three seconds. Both the yellow and green LEDs will then light for 

15 seconds.

Within 10 seconds of pressing the programming button, place pressure on the 

sensor pad and then release. The yellow LED will blink once, confirming the 

monitor and pad are now synchronized and ready for use.

Press and hold the PROGRAM button first, then press and hold the RESET 

button at the same time, for two seconds. The yellow and green LEDs will 

simultaneously flash three times indicating that the pad(s) has 

been unsynchronized.

1

1

1

3

2

2

2

MONITOR SETUP

SYNCHRONIZING ONE CORDLESS SENSOR OR A SECOND 

CORDLESS SENSOR 

(MAKE SURE THE TRANSMITTER IS SETUP PRIOR TO SYNCHRONIZING)

UNSYNCHRONIZING A CORDLESS SENSOR

©2017 Curbell Medical Products, Inc.     MAP1679B

www.curbellmedical.com • 

1-888-222-2188

CSM-BC500 QUICK SETUP GUIDE

NOTE:

 Consult the “Directions For Use” included with monitor for 

complete instructions, troubleshooting, and safety information

1

2

3

TEST THE ALARM:

1. When pressure is placed back on the pad, the green LED will blink two times and then continuously every three seconds indicating the pad is in use.

2. Remove pressure from the pad to activate the alarm. The red LED will then blink (for 10 seconds) and nurse call will alarm.

3. To cancel the alarm, press the monitor’s RESET button and cancel the nurse call alarm at the patient station.

3

1

2

3