Chapter 1: Introduction
WARNING
Use watertight compression cable glands rated IP66 for each cable used.
WARNING
Use silicone to seal the cable gland rated IP66 if more than one cable is use in the
same cable gland.
WARNING
The warranty is void if the product enclosures are not sealed correctly and the
installation does not respect the manufacturer recommendations.
WARNING
Ensure all cables enter through the bottom of the plastic enclosure. Do not make
holes on the top or on the sides of the enclosures. Be careful not to damage the
electronic cards located inside the enclosure when drilling at the bottom of the
enclosure.
NOTICE
The use of flexible tube with water and dust tight connectors at both ends is
acceptable.
Cable support
Support the cables with clips or cable trays whenever possible to avoid damage at the connection points.
Cable routing
NOTICE
Never run low voltage (24V and less) wires like communication wires, inputs or
sensors wires in the same conduit as a High Voltage (Power) wire.
When low voltage cables run parallel to high voltage cables (120/230/380 VAC or 24 VDC), place them at
a distance of at least 300 mm (12 inches) from each other to avoid electromagnetic interference
If low voltage cables cross high voltage cables, ensure they cross at an angle of 90° to minimize electro-
magnetic interference.
895-00700
Bird scale
11
Summary of Contents for 890-00610
Page 4: ...NOTES 4 895 00700 Bird scale...
Page 14: ...NOTES 14 895 00700 Bird scale...
Page 20: ...NOTES 20 895 00700 Bird scale...
Page 22: ...NOTES 22 895 00700 Bird scale...
Page 26: ...NOTES 26 895 00700 Bird scale...
Page 28: ...NOTES 28 895 00700 Bird scale...
Page 34: ...NOTES 34 895 00700 Bird scale...
Page 38: ...NOTES 38 895 00700 Bird scale...
Page 40: ...NOTES 40 895 00700 Bird scale...
Page 42: ...NOTES 42 895 00700 Bird scale...
Page 44: ...NOTES 44 895 00700 Bird scale...
Page 48: ...NOTES 48 895 00700 Bird scale...
Page 50: ...NOTES 50 895 00700 Bird scale...
Page 54: ...NOTES 54 895 00700 Bird scale...