Care and maintenance
26
EN-US
Cubusan CP120
Fault
Cause
Elimination
3) No air coming out of the air
outlet, fan noise present and de-
vice lighting working
Air intake blocked
Check that all packaging materials have been
removed and that there are no foreign bodies
at/in the air inlet (at the bottom, all round the
device)
Air outlet blocked
Check that all packaging materials have been
removed and that there are no foreign bodies
at/in the air outlet (at the top, all round the de-
vice)
Device is being operated in a location
that does not comply with the specifi-
cations
Position the device in a suitable location
chapt. Site selection, page 24)
Fan defective
Have the device checked/repaired by the
manufacturer
4) Odor in room
Insufficient supply of fresh air to the
room
Increase the fresh air supply to the ventilation
system or manually ventilate the room more
regularly
(see chapt. Site selection, page 24)
If this does not resolve the problem, have the
device checked by the manufacturer.