9. Settings (only by trained specialists)
The housing may only be opened by trained specialists!
Standard settings:
Usually the
3
IR remote control is supplied fully assembled and equipped ex
works. This means that no further changes need to be made to the hardware
at this point. If there are several (up to 4) bed controls in one room, it must be
ensured that each remote control controls its own bed. For this, the addresses
of the respective
3
IR remote controls and
BIF8plus
must be set:
•
3
Change the adress of the remote control
Press
and with the appropriate adress (red number).
• Change BIF8plus adress with the
3
IR remote control
Press
and again and confirm with .
Address/button 1 is used as the default address in the factory settings.
Individual settings with BIF software:
•
1
Connect BIF8plus to power supply
Supply the
BIF8plus
with power
(220V)
.
• Connect BIF8plus to PC
Connect the
BIF8plus
to the PC via USB.
• Start software
Then start the Universal Configurator software.
• Assign functions
Now you can assign the functions to the appropriate relays by clicking on
them. Confirm your final entry with
1
Set
.
Summary of Contents for BIF8plus